What is the measurement rate of the DCT 531 digital pressure transmitter?

Measurement Rate of the DCT 531 Digital Pressure Transmitter

The DCT 531 digital pressure transmitter offers highly accurate pressure measurement with the following key performance metrics:

  • Power-On Delay: The DCT 531 has a power-on delay time of 500 milliseconds. This means that after powering on, the device will take approximately 0.5 seconds before it is ready to provide accurate measurements.
  • Measurement Rate: The typical measurement rate of the DCT 531 is 500 Hz, which translates to a measurement every 2 milliseconds (0.002 seconds). This high-frequency rate ensures that the transmitter captures rapid pressure changes in dynamic environments.
  • RS485 Communication and Baud Rate Impact:
    The actual number of measured values per second depends on the baud rate of the RS485 communication setup. Below is a summary of the measurement rates at different baud rates: Baud Rate Measured Values / Second 115 kBit/s Approx. 140 38.4 kBit/s Approx. 100 19.2 kBit/s Approx. 65 9.6 kBit/s Approx. 35

This variability in measured values is due to the different data transmission speeds allowed by each baud rate, affecting how quickly data can be communicated and processed.

Wiring diagram for convenience: